In addition to confounding mass-based wear measurements in serum-lubricated hip simulator experiments, fluid absorption by the acetabular cups may simultaneously modify the wear resistance of the ultra-high molecular weight polyethylene (UHMWPE) from which they are composed. To decouple the fluid absorption and wear processes enabling clearer investigation of this effect, absorption was first imposed during an initial stage where UHMWPE was exposed to pressurized (10MPa) fluid. This was followed by a second stage, where resultant wear behavior was assessed by a multidirectional pin-on-flat technique that, though still providing a serum-lubricating environment, does not promote the simultaneous fluid absorption occurring in hip simulator testing. Both unirradiated and highly-crosslinked UHMWPE were investigated, each with both bovine calf serum and water soaking exposures of duration to 129 days. The pressurized soaking of a highly-crosslinked UHMWPE decreased its wear resistance, causing an increase in wear rate by approximately 50% during subsequent serum-lubricated multidirectional pin-on-flat sliding tests as compared to non-soaked material. The magnitude of this effect did not appear to depend on whether the soaking fluid was water or serum, nor did it appear to depend on soak time provided it was at least of a 14-day duration during which more rapid transient fluid absorption occurs. Such soaking did not produce as pronounced an effect on unirradiated UHMWPE, at its lack of wear resistance likely causes the absorption-affected surface region to the completed removed within the earliest stages of sliding contact.
Skip Nav Destination
ASME/STLE 2004 International Joint Tribology Conference
October 24–27, 2004
Long Beach, California, USA
Conference Sponsors:
- Tribology Division
ISBN:
0-7918-4181-2
PROCEEDINGS PAPER
Effect of Water and Serum Absorption on Wear of Unirradiated and Crosslinked UHMWPE Orthopaedic Bearing Materials
Kathryn Laraia,
Kathryn Laraia
Rensselaer Polytechnic Institute, Troy, NY
Search for other works by this author on:
Nicholas Leone,
Nicholas Leone
Rensselaer Polytechnic Institute, Troy, NY
Search for other works by this author on:
Rebecca MacDonald,
Rebecca MacDonald
Rensselaer Polytechnic Institute, Troy, NY
Search for other works by this author on:
Thierry Blanchet
Thierry Blanchet
Rensselaer Polytechnic Institute, Troy, NY
Search for other works by this author on:
Kathryn Laraia
Rensselaer Polytechnic Institute, Troy, NY
Nicholas Leone
Rensselaer Polytechnic Institute, Troy, NY
Rebecca MacDonald
Rensselaer Polytechnic Institute, Troy, NY
Thierry Blanchet
Rensselaer Polytechnic Institute, Troy, NY
Paper No:
TRIB2004-64008, pp. 1689-1694; 6 pages
Published Online:
December 22, 2008
Citation
Laraia, K, Leone, N, MacDonald, R, & Blanchet, T. "Effect of Water and Serum Absorption on Wear of Unirradiated and Crosslinked UHMWPE Orthopaedic Bearing Materials." Proceedings of the ASME/STLE 2004 International Joint Tribology Conference. ASME/STLE 2004 International Joint Tribology Conference, Parts A and B. Long Beach, California, USA. October 24–27, 2004. pp. 1689-1694. ASME. https://doi.org/10.1115/TRIB2004-64008
Download citation file:
3
Views
0
Citations
Related Proceedings Papers
Related Articles
Serum Lubricant Absorption by UHMWPE Orthopaedic Bearing Implants
J. Tribol (January,2002)
HIP Action
Mechanical Engineering (December,2004)
Effect of Contact Area on the Wear and Friction of UHMWPE in Circular Translation Pin-on-Disk Tests
J. Tribol (November,2017)
Related Chapters
Materials
Design and Application of the Worm Gear
Concluding Remarks
Design of Mechanical Bearings in Cardiac Assist Devices
Unbalance
Fundamentals of Rotating Machinery Diagnostics