Grid-Induced Numerical Errors for Shear Stresses and Essential Flow Variables in a Ventricular Assist Device: Crucial for Blood Damage Prediction?

[+] Author and Article Information
Lucas Konnigk

Institute of Turbomachinery,
Faculty of Mechanical Engineering and
Marine Technology,
University of Rostock,
Albert-Einstein-Straße 2,
Rostock 18055, Germany
e-mail: lucas.konnigk@uni-rostock.de

Benjamin Torner

Institute of Turbomachinery,
Faculty of Mechanical Engineering and
Marine Technology,
University of Rostock,
Albert-Einstein-Straße 2,
Rostock 18055, Germany
e-mail: benjamin.torner@uni-rostock.de

Sebastian Hallier

Institute of Turbomachinery,
Faculty of Mechanical Engineering and
Marine Technology,
University of Rostock,
Albert-Einstein-Straße 2,
Rostock 18055, Germany
e-mail: sebastian.hallier@uni-rostock.de

Matthias Witte

Institute of Turbomachinery,
Faculty of Mechanical Engineering and
Marine Technology,
University of Rostock,
Albert-Einstein-Straße 2,
Rostock 18055, Germany
e-mail: matthias.witte@uni-rostock.de

Frank-Hendrik Wurm

Institute of Turbomachinery,
Faculty of Mechanical Engineering and
Marine Technology,
University of Rostock,
Albert-Einstein-Straße 2,
Rostock 18055, Germany
e-mail: hendrik.wurm@uni-rostock.de

1Corresponding author.

Manuscript received December 15, 2017; final manuscript received February 19, 2019; published online April 1, 2019. Assoc. Editor: Marc Horner.

J. Verif. Valid. Uncert 3(4), 041002 (Apr 01, 2019) (10 pages) Paper No: VVUQ-17-1037; doi: 10.1115/1.4042989 History: Received December 15, 2017; Revised February 19, 2019

Adverse events due to flow-induced blood damage remain a serious problem for blood pumps as cardiac support systems. The numerical prediction of blood damage via computational fluid dynamics (CFD) is a helpful tool for the design and optimization of reliable pumps. Blood damage prediction models primarily are based on the acting shear stresses, which are calculated by solving the Navier–Stokes equations on computational grids. The purpose of this paper is to analyze the influence of the spatial discretization and the associated discretization error on the shear stress calculation in a blood pump in comparison to other important flow quantities like the pressure head of the pump. Therefore, CFD analysis using seven unsteady Reynolds-averaged Navier–Stokes (URANS) simulations was performed. Two simple stress calculation indicators were applied to estimate the influence of the discretization on the results using an approach to calculate numerical uncertainties, which indicates discretization errors. For the finest grid with 19 × 106 elements, numerical uncertainties up to 20% for shear stresses were determined, while the pressure heads show smaller uncertainties with a maximum of 4.8%. No grid-independent solution for velocity gradient-dependent variables could be obtained on a grid size that is comparable to mesh sizes in state-of-the-art blood pump studies. It can be concluded that the grid size has a major influence on the shear stress calculation, and therefore, the potential blood damage prediction, and that the quantification of this error should always be taken into account.

FIGURES IN THIS ARTICLE
<>
Copyright © 2018 by ASME
Your Session has timed out. Please sign back in to continue.

References

Graefe, R. , Henseler, A. , and Steinseifer, U. , 2016, “ Multivariate Assessment of the Effect of Pump Design and Pump Gap Design Parameters on Blood Trauma,” Artif. Organs, 40(6), pp. 568–576. [CrossRef] [PubMed]
Buja, L. M. , and Butany, J. , eds., 2016, Cardiovascular Pathology, 4th ed., Elsevier, London, UK.
Koliopoulou, A. , McKellar, S. H. , Rondina, M. , and Selzman, C. H. , 2016, “ Bleeding and Thrombosis in Chronic Ventricular Assist Device Therapy: Focus on Platelets,” Curr. Opin. Cardiol., 31(3), pp. 299–307. [CrossRef] [PubMed]
Leone, O. , 2017, The Pathology of Cardiac Transplantation: A Clinical and Pathological Perspective, Springer, Cham, Switzerland.
Pagani, F. D. , Miller, L. W. , Russell, S. D. , Aaronson, K. D. , John, R. , Boyle, A. J. , Conte, J. V. , Bogaev, R. C. , MacGillivray, T. E. , Naka, Y. , Mancini, D. , Massey, H. T. , Chen, L. , Klodell, C. T. , Aranda, J. M. , Moazami, N. , Ewald, G. A. , Farrar, D. J. , and Frazier, O. H. , 2009, “ Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist Device,” J. Am. Coll. Cardiol., 54(4), pp. 312–321. [CrossRef] [PubMed]
Olia, S. E. , Maul, T. M. , Antaki, J. F. , and Kameneva, M. V. , 2016, “ Mechanical Blood Trauma in Assisted Circulation: Sublethal RBC Damage Preceding Hemolysis,” Int. J. Artif. Organs, 39(4), pp. 150–159. [CrossRef] [PubMed]
Thamsen, B. , Granneger, M. , and Kretzscher, U. , 2016, “ Blood Damage in Ventricular Assist Devices,” Int. J. Artif Organs, 39(4), pp. 147–149. [CrossRef] [PubMed]
Haeck, M. L. A. , Hoogslag, G. E. , Rodrigo, S. F. , Atsma, D. E. , Klautz, R. J. , van der Wall, E. E. , Schalij, M. J. , and Verwey, H. F. , 2012, “ Treatment Options in End-Stage Heart Failure: Where to Go From Here?,” Neth. Heart J., 20(4), pp. 167–175. [CrossRef] [PubMed]
Taskin, M. E. , Fraser, K. H. , Zhang, T. , Gellman, B. , Fleischli, A. , Dasse, K. A. , Griffith, B. P. , and Wu, Z. J. , and., 2010, “ Computational Characterization of Flow and Hemolytic Performance of the UltraMag Blood Pump for Circulatory Support,” Artif. Organs, 34(12), pp. 1099–1113. [CrossRef] [PubMed]
Fraser, K. H. , Zhang, T. , Taskin, M. E. , Griffith, B. P. , and Wu, Z. J. , 2012, “ A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index,” ASME J. Biomech. Eng., 134(8), p. 81002. [CrossRef]
Farinas, M.-I. , Garon, A. , and Lacasse, D. , 2006, “ Asymptotically Consistent Numerical Approximation of Hemolysis,” ASME J. Biomech. Eng., 128(5), pp. 688–696.
Arwatz, G. , and Smits, A. J. , 2013, “ A Viscoelastic Model of Shear-Induced Hemolysis in Laminar Flow,” Biorheology, 50(1–2), pp. 45–55. [PubMed]
Goubergrits, L. , and Affeld, K. , 2004, “ Numerical Estimation of Blood Damage in Artificial Organs,” Artif. Organs, 28(5), pp. 499–507. [CrossRef] [PubMed]
Grigioni, M. , Daniele, C. , Morbiducci, U. , D'Avenio, G. , Di Benedetto, G. , and Barbaro, V. , 2004, “ The Power-Law Mathematical Model for Blood Damage Prediction: Analytical Developments and Physical Inconsistencies,” Artif. Organs, 28(5), pp. 467–475. [CrossRef] [PubMed]
Hallier, S. , Torner, B. , Kumar, J. , and Wurm, F.-H. , 2016, “ Miniaturization of an Implantable Pump for Heart Support,” International Symposium on Transport Phenomena and Dynamics of Rotating Machinery (ISROMAC '16), Honolulu, HI, Apr. 10–15. http://isromac-isimet.univ-lille1.fr/upload_dir/finalpaper/264.finalpaper.pdf
Throckmorton, A. L. , Lim, D. S. , McCulloch, M. A. , Jiang, W. , Song, X. , Allaire, P. E. , Wood, H. G. , and Olsen, D. B. , 2005, “ Computational Design and Experimental Performance Testing of an Axial-Flow Pediatric Ventricular Assist Device,” ASAIO J., 51(5), pp. 629–635. [CrossRef] [PubMed]
Kido, K. , Hoshi, H. , Watanabe, N. , Kataoka, H. , Ohuchi, K. , Asama, J. , Shinshi, T. , Yoshikawa, M. , and Takatani, S. , 2006, “ Computational Fluid Dynamics Analysis of the Pediatric Tiny Centrifugal Blood Pump (TinyPump),” Artif. Organs, 30(5), pp. 392–399. [CrossRef] [PubMed]
Chen, Z. , Jena, S. K. , Giridharan, G. A. , Koenig, S. C. , Slaughter, M. S. , Griffith, B. P. , and Wu, Z. J. , 2018, “ Flow Features and Device-Induced Blood Trauma in CF-VADs Under a Pulsatile Blood Flow Condition: A CFD Comparative Study,” Int. J. Numer. Methods Biomed. Eng., 34(2), (epub).
Schüle, C. Y. , Thamsen, B. , Blumel, B. , Lommel, M. , Karakaya, T. , Paschereit, C. O. , Affeld, K. , and Kertzscher, U. , 2016, “ Experimental and Numerical Investigation of an Axial Rotary Blood Pump,” Artif. Organs, 40(11), pp. 192–202. [CrossRef]
Fraser, K. H. , Taskin, M. E. , Griffith, B. P. , and Wu, Z. J. , 2011, “ The Use of Computational Fluid Dynamics in the Development of Ventricular Assist Devices,” Med. Eng. Phys., 33(3), pp. 263–280. [CrossRef] [PubMed]
ANSYS, Inc., 2017, ANSYS CFX–Modeling Guide 18.0: Chap. 15.3.2, ANSYS, Canonsburg, PA.
ANSYS, Inc., 2017, ANSYS CFX–Theory Guide 18.0: Chap. 11.1, ANSYS, Canonsburg, PA.
ANSYS, Inc., 2017, ANSYS CFX–Theory Guide 18.0: Chap. 11.1.1.4.5, ANSYS, Canonsburg, PA.
Huang, Z. J. , Merkle, C. L. , Abdallah, S. , and Tarbell, J. M. , 1994, “ Numerical Simulation of Unsteady Laminar Flow Through a Tilting Disk Heart Valve: Prediction of Vortex Shedding,” J. Biomech., 27(4), pp. 391–402. [CrossRef] [PubMed]
Menter, F. R. , 1994, “ Two-Equation Eddy-Viscosity Turbulence Models for Engineering Applications,” AIAA J., 32(8), pp. 1598–1605. [CrossRef]
Smirnov, P. E. , and Menter, F. R. , 2009, “ Sensitization of the SST Turbulence Model to Rotation and Curvature by Applying the Spalart–Shur Correction Term,” ASME J. Turbomach., 131(4), p. 041010.
Menter, F. R. , Langtry, R. , and Völker, S. , 2006, “ Transition Modelling for General Purpose CFD Codes,” Flow Turbul. Combust, 77(1–4), pp. 277–303. [CrossRef]
ANSYS Inc., 2017, ANSYS CFX–Theory Guide 18.0: Chap. 11.1.1.4.3, ANSYS, Canonsburg, PA.
Wurzinger, L. J. , Opitz, R. , and Eckstein, H. , 1986, “ Mechanical Bloodtrauma. An Overview,” Angeiologie, 38(3), pp. 81–97.
Giersiepen, M. , Wurzinger, L. J. , Opitz, R. , and Reul, H. , 1990, “ Estimation of Shear Stress-Related Blood Damage in Heart Valve Prostheses—In Vitro Comparison of 25 Aortic Valves,” Int. J. Artif. Organs, 13(5), pp. 300–306. [CrossRef] [PubMed]
Garon, A. , and Farinas, M.-I. , 2004, “ Fast Three-Dimensional Numerical Hemolysis Approximation,” Artif. Organs, 28(11), pp. 1016–1025. [CrossRef] [PubMed]
Taskin, M. E. , Fraser, K. H. , Zhang, T. , Wu, C. , Griffith, B. P. , and Wu, Z. J. , 2012, “ Evaluation of Eulerian and Lagrangian Models for Hemolysis Estimation,” ASAIO J., 58(4), pp. 363–372. [CrossRef] [PubMed]
Yu, H. , Engel, S. , Janiga, G. , and Thévenin, D. , 2017, “ A Review of Hemolysis Prediction Models for Computational Fluid Dynamics,” Artif. Organs, 41(7), pp. 603–621. [CrossRef] [PubMed]
Casa, L. D. C. , and Ku, D. N. , 2017, “ Thrombus Formation at High Shear Rates,” Annu. Rev. Biomed. Eng., 19(1), pp. 415–433. [CrossRef] [PubMed]
Hellums, J. D. , 1994, “ Biorheology in Thrombosis Research,” Annu. Biomed. Eng., 22(5), pp. 445–455. [CrossRef]
Thamsen, B. , Blumel, B. , Schaller, J. , Paschereit, C. O. , Affeld, K. , Goubergrits, L. , and Kertzscher, U. , 2015, “ Numerical Analysis of Blood Damage Potential of the HeartMate II and HeartWare HVAD Rotary Blood Pumps,” Artif. Organs, 39(8), pp. 651–659. [CrossRef] [PubMed]
Eça, L. R. C. , and Hoekstra, M. , 2014, “ A Procedure for the Estimation of the Numerical Uncertainty of CFD Calculations Based on Grid Refinement Studies,” J. Comput. Phys., 262, pp. 104–130. [CrossRef]
Guelich, J. F. , 2010, Kreiselpumpen: Handbuch Für Entwicklung, Anlagenplanung Und Betrieb, 3rd ed., Springer, Berlin/Heidelberg.
Roache, P. J. , 1998, Verification and Validation in Computational Science and Engineering, Hermosa Publishers, Albuquerque, NM.
Roache, P. J. , 1997, “ Quantification of Uncertainty in Computational Fluid Dynamics,” Annu. Rev. Fluid Mech., 29, pp. 126–160. [CrossRef]
Stewart, S. F. C. , Paterson, E. G. , Burgreen, G. W. , Hariharan, P. , Giarra, M. , Reddy, V. , Day, S. W. , Manning, K. B. , Deutsch, S. , Berman, M. R. , Myers, M. R. , and Malinauskas, R. A. , 2012, “ Assessment of CFD Performance in Simulations of an Idealized Medical Device: Results of FDA's First Computational Interlaboratory Study,” Cardiovasc. Eng. Tech., 3(2), pp. 139–160. [CrossRef]
Moser, R. D. , Kim, J. , and Mansour, N. N. , 1999, “ Direct Numerical Simulation of Turbulent Channel Flow Up to Reτ = 590,” Phys. Fluids, 11(4), pp. 943–945. [CrossRef]
Menter, F. R. , 2009, “ Review of the Shear-Stress Transport Turbulence Model Experience From an Industrial Perspective,” Int. J. Comput. Fluid Dyn., 23(4), pp. 305–316. [CrossRef]
Torner, B. , Konnigk, L. , Hallier, S. , Kumar, J. , Witte, M. , and Wurm, F.-H. , 2018, “ Large Eddy Simulation in a Rotary Blood Pump: Viscous Shear Stress Computation and Comparison With Unsteady Reynolds-Averaged Navier-Stokes Simulation,” Int. J. Artif. Organs, 41(11), pp. 752–763. [CrossRef] [PubMed]

Figures

Grahic Jump Location
Fig. 1

Numerical model of the analyzed VAD with a two-bladed rotor, an inlet guide vane with five blades, and a three-bladed outlet guide vane

Grahic Jump Location
Fig. 2

Left: section of the computational grid in the rotor for finest and coarsest meshes; right: detailed view of the mesh in gap between rotor and casing

Grahic Jump Location
Fig. 3

Section of FDA's benchmark testcase nozzle with sudden expansion (edited from Ref. [41]). Red lines represent locations where the uncertainty estimation was conducted with time-averaged velocity results.

Grahic Jump Location
Fig. 4

(a) Centerline velocity and (b) velocity profile. Black solid line: experimental data with uncertainties. Dashed line: results from CFD literature results. Red circles: Our results obtained on the finest grid with estimated uncertainties. Experimental and CFD data from group which also used k−ω model and contributed to the round robin study are obtained from the website.2

Grahic Jump Location
Fig. 5

(a) Comparison of time-averaged velocity profile from bottom wall to channel center between literature data from Moser et al. [42] (black line) and our DNS (red rots). The calculated uncertainties are barely visible at most of the data points and (b) comparison between exact solution (black line) and solution obtained on the coarsest grid (red dots) with calculated uncertainties.

Grahic Jump Location
Fig. 6

Pressure heads via the rotor (top). The line shows the fitHr¯=109.3−3.66×107hi2. Pressure heads via the pump (bottom). The line shows the fitHp¯=74.5−7.49×107hi2. The error bars mark the numerical uncertainties (deviations in percent).

Grahic Jump Location
Fig. 7

Global volume integral of shear stresses over the pump. The line shows the fit Iτ=1.14×10−4−0.017×hi0.89. The error bars mark the numerical uncertainties (deviations in percent).

Grahic Jump Location
Fig. 8

Upper left: volume in pump which exceeds 9 Pa. The line shows the fitVτ¯>9 Pa=1.98×10−6−7.78hi2. Upper right: volume in pump which exceeds 50 Pa. The line shows the fitVτ¯>50 Pa=2.61×10−7−1.90×10−3hi1.25. Bottom: volume in pump which exceeds 150 Pa. The line shows the fitVτ¯>150 Pa=3.37×10−8−0.48hi2.08. The error bars mark the numerical uncertainties (deviations in percent).

Grahic Jump Location
Fig. 9

Time-averaged wall shear stresses on the surface of the rotor (top) and leading edge (bottom) between coarsest (UR-1) and finest (UR-7) mesh

Grahic Jump Location
Fig. 10

Volumes with shear stresses over 150 Pa in the rotor of the VAD between finest (UR-7) and coarsest (UR-1) mesh

Tables

Errata

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In