Tekscan pressure sensors are used in biomechanics research to measure joint contact loads. While the overall accuracy of these sensors has been reported previously, the effects of different calibration algorithms on sensor accuracy have not been compared. The objectives of this validation study were to determine the most appropriate calibration method supplied in the Tekscan program software and to compare its accuracy to the accuracy obtained with two user-defined calibration protocols. We evaluated the calibration accuracies for test loads within the low range, high range, and full range of the sensor. Our experimental setup used materials representing those found in standard prosthetic joints, i.e., metal against plastic. The Tekscan power calibration was the most accurate of the algorithms provided with the system software, with an overall rms error of 2.7% of the tested sensor range, whereas the linear calibrations resulted in an overall rms error of up to 24% of the tested range. The user-defined ten-point cubic calibration was almost five times more accurate, on average, than the power calibration over the full range, with an overall rms error of 0.6% of the tested range. The user-defined three-point quadratic calibration was almost twice as accurate as the Tekscan power calibration, but was sensitive to the calibration loads used. We recommend that investigators design their own calibration curves not only to improve accuracy but also to understand the range(s) of highest error and to choose the optimal points within the expected sensing range for calibration. Since output and sensor nonlinearity depend on the experimental protocol (sensor type, interface shape and materials, sensor range in use, loading method, etc.), sensor behavior should be investigated for each different application.
Skip Nav Destination
e-mail: canglin@ucalgary.ca
Article navigation
March 2009
Technical Briefs
Effect of Calibration Method on Tekscan Sensor Accuracy
Jill M. Brimacombe,
Jill M. Brimacombe
Department of Mechanical Engineering,
University of British Columbia
, 6250 Applied Science Lane, Vancouver, BC, V6R 2L7, Canada
Search for other works by this author on:
David R. Wilson,
David R. Wilson
Division of Orthopaedic Engineering Research, Department of Orthopaedics, and Vancouver Coastal Health Research Institute,
University of British Columbia
, 500-828 West 10th Avenue, Vancouver, BC, V6K 1L8, Canada
Search for other works by this author on:
Antony J. Hodgson,
Antony J. Hodgson
Department of Mechanical Engineering,
University of British Columbia
, 6250 Applied Science Lane, Vancouver, BC, V6R 2L7, Canada
Search for other works by this author on:
Karen C. T. Ho,
Karen C. T. Ho
Centre for Bioengineering Research and Education, and Department of Civil Engineering,
University of Calgary
, 2500 University Drive NW, Calgary, AB, T2N 1N4, Canada
Search for other works by this author on:
Carolyn Anglin
Carolyn Anglin
Centre for Bioengineering Research and Education, and Department of Civil Engineering,
e-mail: canglin@ucalgary.ca
University of Calgary
, 2500 University Drive NW, Calgary, AB, T2N 1N4, Canada
Search for other works by this author on:
Jill M. Brimacombe
Department of Mechanical Engineering,
University of British Columbia
, 6250 Applied Science Lane, Vancouver, BC, V6R 2L7, Canada
David R. Wilson
Division of Orthopaedic Engineering Research, Department of Orthopaedics, and Vancouver Coastal Health Research Institute,
University of British Columbia
, 500-828 West 10th Avenue, Vancouver, BC, V6K 1L8, Canada
Antony J. Hodgson
Department of Mechanical Engineering,
University of British Columbia
, 6250 Applied Science Lane, Vancouver, BC, V6R 2L7, Canada
Karen C. T. Ho
Centre for Bioengineering Research and Education, and Department of Civil Engineering,
University of Calgary
, 2500 University Drive NW, Calgary, AB, T2N 1N4, Canada
Carolyn Anglin
Centre for Bioengineering Research and Education, and Department of Civil Engineering,
University of Calgary
, 2500 University Drive NW, Calgary, AB, T2N 1N4, Canadae-mail: canglin@ucalgary.ca
J Biomech Eng. Mar 2009, 131(3): 034503 (4 pages)
Published Online: December 23, 2008
Article history
Received:
October 7, 2007
Revised:
July 18, 2008
Published:
December 23, 2008
Citation
Brimacombe, J. M., Wilson, D. R., Hodgson, A. J., Ho, K. C. T., and Anglin, C. (December 23, 2008). "Effect of Calibration Method on Tekscan Sensor Accuracy." ASME. J Biomech Eng. March 2009; 131(3): 034503. https://doi.org/10.1115/1.3005165
Download citation file:
Get Email Alerts
Related Articles
Design and Implementation of Noncontact Measurement Device for Irregular Dental Osseointegration Detection
J. Med. Devices (June,2011)
Influence of Force Field Direction on Pressure Sensors Calibrated at
Up to 12,000 g
J. Eng. Gas Turbines Power (November,2008)
Designing an Optical Bendloss Sensor for Clinical ForcMeasurement
J. Med. Devices (June,2009)
The Momedemameter: An Affordable Pre-Eclampsia Detector for Low Resource Environments
J. Med. Devices (March,2009)
Related Proceedings Papers
Related Chapters
Analysis of ASTM Round-Robin Results: Ae Sensors, Systems, Brands, and Calibration Rods
Acoustic Emission Testing of Aerial Devices and Associated Equipment Used in the Utility Industries
4-Bit Pipeline ADC for Monolithic Active Pixel Sensors
International Conference on Advanced Computer Theory and Engineering (ICACTE 2009)
Getting Ready and Staying Ready: Preparation, Verification, and Maintenance of Apparatus
The Practice of Flash Point Determination: A Laboratory Resource