In this study, a new nitinol based fixation device was investigated for use in repairing severed digital flexor tendons. The device, composed of superelastic nitinol, is tubular in shape with inward facing tines for gripping tissue. Its cellular structure was designed such that it has a large effective Poisson's ratio, which facilitates a “finger trap” effect. This allows for reduced tendon compression during a resting state (to permit vascular perfusion) and increased compression during loading (to drive the tines into the tissue for gripping). To test the feasibility of using this device for flexor tendon repair, it was tested on cadaver flexor digitorum profundus tendons. The tendons were excised, cut in the region corresponding to a zone II laceration, and repaired using the device. The device was easy to install and did not prevent the tendon from bending. Constant strain rate tensile testing revealed a mean tensile strength of 57.6 ± 7.7 N, with a force of 53.2 ± 7.8 N at a 2 mm gap. This exceeds the suggested primary repair strength of 45 N, which has been proposed as the necessary strength for enabling early mobilization. Although considerable future studies will be needed to determine the suitability of the new repair device for clinical use, this study demonstrates the feasibility of utilizing a tubular, nitinol repair device for flexor tendon fixation.
Skip Nav Destination
Article navigation
January 2014
Technical Briefs
A Nitinol Based Flexor Tendon Fixation Device: Gapping and Tensile Strength Measurements in Cadaver Flexor Tendon
Shawn P. Reese,
Shawn P. Reese
1
CoNextions Inc.,
Salt Lake City, UT 84104;
Department of Bioengineering,
e-mail: s.reese@conextionsmed.com
Salt Lake City, UT 84104;
Department of Bioengineering,
The University of Utah
,Salt Lake City, UT 84112
e-mail: s.reese@conextionsmed.com
1Corresponding author.
Search for other works by this author on:
Erik N. Kubiak
Erik N. Kubiak
CoNextions Inc.,
Salt Lake City, UT 84104;
Division of Orthopaedic Trauma,
Department of Orthopaedic Surgery,
Salt Lake City, UT 84104;
Division of Orthopaedic Trauma,
Department of Orthopaedic Surgery,
The University of Utah
,Salt Lake City, UT 84112
Search for other works by this author on:
Shawn P. Reese
CoNextions Inc.,
Salt Lake City, UT 84104;
Department of Bioengineering,
e-mail: s.reese@conextionsmed.com
Salt Lake City, UT 84104;
Department of Bioengineering,
The University of Utah
,Salt Lake City, UT 84112
e-mail: s.reese@conextionsmed.com
Erik N. Kubiak
CoNextions Inc.,
Salt Lake City, UT 84104;
Division of Orthopaedic Trauma,
Department of Orthopaedic Surgery,
Salt Lake City, UT 84104;
Division of Orthopaedic Trauma,
Department of Orthopaedic Surgery,
The University of Utah
,Salt Lake City, UT 84112
1Corresponding author.
Contributed by the Bioengineering Division of ASME for publication in the JOURNAL OF BIOMECHANICAL ENGINEERING. Manuscript received May 8, 2013; final manuscript received September 30, 2013; accepted manuscript posted October 19, 2013; published online December 3, 2013. Assoc. Editor: Zong-Ming Li.
J Biomech Eng. Jan 2014, 136(1): 014501 (5 pages)
Published Online: December 3, 2013
Article history
Received:
May 8, 2013
Revision Received:
September 30, 2013
Accepted:
October 19, 2013
Citation
Reese, S. P., and Kubiak, E. N. (December 3, 2013). "A Nitinol Based Flexor Tendon Fixation Device: Gapping and Tensile Strength Measurements in Cadaver Flexor Tendon." ASME. J Biomech Eng. January 2014; 136(1): 014501. https://doi.org/10.1115/1.4025779
Download citation file:
Get Email Alerts
Cited By
Related Articles
Effects of Static Stress on the Mechanical Properties of Cultured Collagen Fascicles From the Rabbit Patellar Tendon
J Biomech Eng (February,2002)
Effects of Cyclic Stress on the Mechanical Properties of Cultured Collagen Fascicles from the Rabbit Patellar Tendon
J Biomech Eng (December,2003)
Finite Element Analysis of the Implantation of a Self-Expanding Stent: Impact of Lesion Calcification
J. Med. Devices (June,2012)
Related Proceedings Papers
Related Chapters
IWE and IWL
Online Companion Guide to the ASME Boiler and Pressure Vessel Codes
Appendix F—Tendon End Anchor Testing
Commentary on Article CC-3000 Design
Creating and Eliminating Workplace Hazards by Design
An Instructional Aid For Occupational Safety and Health in Mechanical Engineering Design