Abstract
Our group has developed a new nitinol endoluminal self-expandable device for microvascular anastomosis. It attaches to each vessel ending with opposite directed microspikes and reaches complete expansion at body temperature, using the nitinol shape memory capacity. The main purpose of this first in vivo trial is to evaluate the mechanical viability of the device and its immediate and early functionality.
A recuperation study with seven New Zealand White rabbits was designed. A 1,96mm outer diameter prototype of the new device was placed on the right femoral artery of each rabbit. Each anastomosis was reassessed on the seventh postoperative day to reevaluate the device function.
The average anastomosis time with the new device was 18 min and 45 seg. It could be easily placed in all the cases with an average of 1,14 (1) complementary stitches needed to achieve a sealed anastomosis. Patency test was positive for all the cases on the immediate assessment. On the one week revision surgery, patency test was negative for the seven rabbits due to blood clot formation inside the device.
The new device that we have developed is simple to use and shows correct immediate functionality. On the early assessment, the presence of a foreign body in the endoluminal space caused blood clot formation. We speculate that a heparin eluting version of the device could avoid thrombosis formation. We consider that the results obtained can be valuable for other endoluminal sutureless devices.